P a g e
Alarm Settings
Click the
Alarm Settings button to edit the alarm settings from the Device Parameters tab.
Reading Rate
Check the
Reading Rate box to choose the reading rate from the Device Parameters tab.
Wrap Around
Check the
Wrap Around box to choose the wrap around memory feature from the Device
Parameters tab.
Device Menu: Start Device
Select
Start Device from the Device Menu to display the following window:
NOTE: Starting the device will erase all readings currently stored in its memory.
The
Start Device window allows the start time and reading rate to be set. The Start Method setting
may be used, along with the
Start Now radio button, to start the data logger immediately.
Alternatively, the user can select the
Delay Start option to delay the start of data collection. The start
time may be delayed up to six months from the current time. The
Manual Start option can be
selected to start the data logger by pushing the pushbutton on the device.
The push button must be
pushed down and held for at least 3 seconds.