12

Calibrating-level

In the calibration menu (

CAL) the measured value can be adjusted.

g

The measured value correction (

CAL) is accessible only, if ConF / InP/ Corr = 1 or 2 was

selected.

Two methods are available

Offset - correction

2-point - correction

g

The

InL.x and InH.x values are presented with one digit. As reference for the correcting calculation the

full resolution is used.

g

Deleting the correction values is done fastest by switching off the measured value correction

Corr = 0

or setting the scaling parameters to a linear behaviour.

+

The values

InL.x and InH.x show the real measured value. The output values OuL.x and OuH.x start

with the preset value.