Key Operation

Display

Turn Power Off

PEAK

ZERO

ZERO

MEM

MEM

UNIT

1. Hold down PEAK and UNIT.
2. Press POWER.

PEAK

ZERO

ZERO

MEM

MEM

UNIT

3. Release POWER.

4.4 Tracking

The FGV-XY uses a load cell and strain gauge as its force sensor. The sensor is affected by environmental changes such as temperature and

humidity. The Tracking function, which is active by default, helps to negate the effects of these changes. Tracking can cause errors when

measuring very small forces. It is recommended that you turn off the Tracking function when measuring these smaller forces. With the gauge's

power off, hold PEAK and UNIT. Press POWER, and keep holding PEAK and UNIT until "oFF" is displayed. Repeat the process to turn Track-

ing back on. Instead of "oFF", "SEt" will be displayed.

Splash Screen

Tracking OFF

Tracking ON

Standard Measuring Mode

Release PEAK and UNIT
after the display shows
"SEt" or "oFF".

4.5 Functions

Key operation

Display

Turn power off

PEAK

ZERO

MEM

MEM

UNIT

1. Hold down ZERO.

2. Press POWER.

PEAK

ZERO

MEM

MEM

UNIT

3. Release POWER.

Splash Screen

Function Mode

Release ZERO after the
display shows "f01".

The following settings are available in Function mode.

Setting

Unit

Options

Default

Measurement Polarity

f01

-0001(minus), 0001 (plus)

0001

Display Update Time

f02

1, 2, 3, 5, 10, 20 (times/second)

3

Auto Power Off

f03

10 (10 minutes), oFF (always on)

10

RS-232C Baud Rate

f04

2400, 4800, 9600, 19200 bps

2400

Response Time

f05

3, 20, 150 (msec)

3

Output Type

f06

ovEr, Hi-Lo

ovEr

1.

2.

1.

2.

FGV

FGV

PEAKPEAK

UNITUNIT

POWER

POWER

PEAKPEAK

UNITUNIT

ZEROZERO

POWER

POWER

ZEROZERO