beautypg.com

4 setting lo limit, 5 comparator display, 6 comparator output – SHIMPO FGV-XY-Series User Manual

Page 15

background image

-14-

5.5.4 Setting LO Limit

The procedure to set the HI limit is used to set the LO limit.
Section 5.5.3 contains this information.

Key

Operation

PEAK

Move to memory mode.

MEM

Register the setting, then move

to standard measuring mode.

Turn power off without any change.

LO limit setting mode

5.5.5 Comparator Display

The following symbols are shown when the comparator function is active.

Measuring value > HI limit

Measuring value < LO limit

Data < HI limit

Data > LO limit

Data > HI limit

Data < LO limit

Comparator Display

Posibilities

5.5.6 Comparator Output

The Comparator function also allows control of an alarm or automated process through its data output port.

When the measured value is greater than the HI limit, the corresponding photo-couple is triggered.

When the measured value is lower than the LO limit, the corresponding photo-couple is triggered.

5.6 Memory

The FGV-XY has three memory modes.

Continuous memory

Allows the recording of up to 1000 data points. The recording starts when you push MEM, and stops when you
push MEM. In addition, the following statistics, gathered between that start and stop, are recorded: positive
maximum value, negative maximum value, positive minimum value, negative minimum value, positive peak
value, negative peak value, average value, standard deviation.

Single memory

Allows the recording of up to 100 data points. Every time MEM is pressed, the value shown on the display is
memorized. If the unit is in Standard mode then the current measured value is recorded. In Peak mode, the unit
records the displayed peak value. In addition, the following statistics are recorded: positive maximum value,
negative maximum value, positive minimum value, negative minimum value, average value, standard deviation.

Standard memory

Allows the recording of up to 50 data points. The recording process is similar to Continuous mode. MEM starts
the recording, and stops the recording. The measured value, when MEM is pushed the second time, is recorded
as a point. The following statistics, gathered between the start and stop, are recorded: positive maximum value,
negative maximum value, positive minimum value, negative minimum value.

In order for the Comparator function to display, the

comparator output option must be set in function f06.

See section 4.5.6 regarding this setting.

The absolute value of the LO limit may never be higher than
the absolute value of the HI limit.

In order for the Comparator function to display, the

comparator output option must be set in function f06.

See section 4.5.6 regarding this setting.

For connection and circuit information, pertaining to

the Comparator output, see section 6.4.

POWER

POWER

PEAKPEAK

MEMORY

MEMORY