-13-

Key operation

Status

PEAK

Move to LO limit setting.

MEM

Register the setting, then move

to standard measuring mode.

Turn power off without any change.

Key operation

Display

Turn power off

PEAK

ZERO

ZERO

MEM

MEM

UNIT

1. Hold down PEAK.

2. Press POWER.

PEAK

ZERO

ZERO

MEM

MEM

UNIT

3. Release POWER.

5.5 Comparator

The Comparator function allows you to create conditions for Go/No Go testing. A high(HI), and low(LO) force limit may be set so that the
FGV-XY's display will show when a measurement is not within the HI and LO settings. In addition, the output port will respond relative to the
display. See section 6.4 regarding the comparator output.
The FGV-XY must be in Comparator mode, and not Overload mode to use the following settings. See section 4.5.6 regarding the output
mode.

5.5.1 Compatrator Overview

5.5.2 Comparator / Memory Settings Mode

Turn POWER off, press PEAK and hold. Press POWER and release (release PEAK after "HI" is displayed).

Splash Screen

PEAK

PEAK

Comparator HI limit

Comparator LO limit

Memory mode

PEAK

Setting

Display

Content

Default factory setting

Comparator HI limit

HI

Value for the HI limit*

0

Comparator LO limit

LO

Value for the LO limit*

0

Memory mode

MEM

Memory mode(Standard, Single, Continuous)

Std

* When HI and LO are set to 00.00 the Comparator function is disabled.

5.5.3 Setting HI Limit

1. Press UNIT to select the HI setting. The digits will start fl ashing.
2. Press UNIT again to change the polarity.
3. Choose the digit to change by pressing ZERO.
4. Press UNIT to increase the selected digit to the desired value.
5. Repeat steps 3 and 4 until the desired value and polarity are correct.
6. Press PEAK. This saves the HI limit and displays the LO limit.
7. Press MEM if you are fi nished with the Comparator and Memory settings.

Key operation

Status

Display

UNIT

Changes polarity

ZERO

Selects the next digit.

UNIT

Changes the selected

digit.

Release PEAK after "HI"
is diaplayed.

The Comparator function's limits can be set outside of the
FGV-XY's working range. This may cause unpredictable
results, and is not reccomended.

1.

2.

The values entered for the HI and LO settings are displayed
in the measurement units last used in Standard mode.

The absolute value of the HI limit may never be lower than
the absolute value of the LO limit.