beautypg.com

Technical data, 3 pressure ranges – KROHNE OPTIBAR PC 5060 C EN User Manual

Page 66

background image

7

TECHNICAL DATA

66

OPTIBAR PC 5060 C

www.krohne.com

04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en

7.3 Pressure ranges

Adjustment

Adjustment

Adjustment

Adjustment

Adjustment range of min/max adjustment in relation to the nominal range:

• Percent value (-10...110%)
• Pressure value (-20...120%)

Adjustment range of the zero / span adjustment in relation to the nominal range

• Zero (-20...+95%)
• Span (-120...+120%)
• Difference between zero and span: max. 120% of the nominal range

Recommended maximum Turn Down (TD): 10:1 (no limit)

Nominal ranges and overload capacity in bar/kPa

Nominal ranges and overload capacity in bar/kPa

Nominal ranges and overload capacity in bar/kPa

Nominal ranges and overload capacity in bar/kPa

This information is provided as an overview and refers to the measuring cell. Limitations due to
the material and design of the process connection as well as the type of pressure selected are
possible. The information given on the nameplate applies.

Nominal range

Max. working pressure
(MWP)

Overload capacity min.
pressure

Max. working pressure

Max. working pressure

Max. working pressure

Max. working pressure

0…+0.025 bar / 0…+2.5 kPa

1

+5 bar / +500 kPa

-0.05 bar / -5 kPa

0…+0.1 bar / 0…+10 kPa

+15 bar / +1500 kPa

-0.2 bar / -20 kPa

0…+0.4 bar / 0…+40 kPa

+30 bar / +3000 kPa

-0.8 bar / -80 kPa

0…+1 bar / 0…+100 kPa

+35 bar / +3500 kPa

-1 bar / -100 kPa

0…+2.5 bar / 0…+250 kPa

+50 bar / +5000 kPa

-1 bar / -100 kPa

0…+10 bar / 0…+1000 kPa

+90 bar / +9000 kPa

-1 bar / -100 kPa

0…+25 bar / 0…+2500 kPa

+130 bar / +13000 kPa

-1 bar / -100 kPa

0…+60 bar / 0…+6000 kPa

+200 bar / +20000 kPa

-1 bar / -100 kPa

0…+100 bar / 0…+10000 kPa

2

+200 bar / +20000 kPa

-1 bar / -100 kPa

-1...0 bar / -100...0 kPa

+35 bar / +3500 kPa

-1 bar / -100 kPa

-1...+1.5 bar / -100...+150 kPa

+40 bar / +4000 kPa

-1 bar / -100 kPa

-1...+10 bar / -100...+1000 kPa

+90 bar / +9000 kPa

-1 bar / -100 kPa

-1...+25 bar / -100...+2500 kPa

+130 bar / +13000 kPa

-1 bar / -100 kPa

-1...+60 bar / -100...+6000 kPa

+200 bar / +20000 kPa

-1 bar / -100 kPa

-1...+100 bar / -100...+10000 kPa

2

+200 bar / +20000 kPa

-1 bar / -100 kPa

-0.05...+0.05 bar / -5...+5 kPa

+15 bar / +1500 kPa

-0.2 bar / -20 kPa

-0.2...+0.2 bar / -20...+20 kPa

+20 bar / +2000 kPa

-0.4 bar / -40 kPa

-0.5...+0.5 bar / -50...+50 kPa

+35 bar / +3500 kPa

-1 bar / -100 kPa

1 Only for measuring cell Ø28 mm
2 Only for measuring cell Ø17.5 mm

.book Page 66 Friday, May 2, 2014 12:49 PM