beautypg.com

Start-up – KROHNE OPTIBAR PC 5060 C EN User Manual

Page 45

background image

START-UP

5

45

OPTIBAR PC 5060 C

www.krohne.com

04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en

5.4.2 Display

5.4.3 Diagnosis

Menu language In this menu item you can set the desired language.

Factory settings: English

German, English, French, Spanish, Portuguese, Italian, Dutch, Russian, Turkish, Polish and Czech

Displayed value

1 and 2

In this menu item you can define how the measured value should be presented on the display.

The factory setting is "Linear percent".

"Level"

Filling height or Process pressure (Adjustment

units), Percent, Scaled, Current output, Linear

percent, Meas. cell temp. and Electronics

temperature

"Process pressure"

Process pressure, Percent, Scaled, Current

output, Linear percent, Meas. cell temp. and

Electronics temperature

Backlight

A backlight on the display is available, which can be turned on or off in this menu. By default, this

function is disabled.

Off, On

Device status

Status signals

Check function, Out of Specification, Maintenance required, Failure

Peak value

The respective minimum and maximum pressure values are stored in the device. Under "Peak

values", these values can be viewed or reset. In addition to the pressure, the minimum and maximum

values of the sensor cell and the electronics temperature is stored. These can be viewed or reset

here.

Peak value

"Process pressure"

Reset peak value

Peak value

"Meas. cell temp."

Reset peak value

Peak value

"Electronics temperature"

Reset peak value

Simulation

In menu item "Simulation", measured values can be simulated via the current output. These are
issued as both analogue and digital (via HART

®

). The simulation is automatically cancelled 60

minutes after the last key stroke.

For applications

"Process pressure"

Process pressure, Percent, Current output, Linear percent, Meas.

cell temp. and Electronics temperature

For applications

"Level"

Filling height or Process pressure (Adjustment units), Percent,

Current output, Linear percent, Meas. cell temp. and Electronics

temperature

.book Page 45 Friday, May 2, 2014 12:49 PM