TECHNICAL DATA

7

65

OPTIBAR PC 5060 C

www.krohne.com

04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en

Inputs and outputs

Output signal

Output signal

Output signal

Output signal

Output signal

4...20 mA / HART

®

version 7.3

3.8...20.5 mA (factory setting acc. to. NAMUR recommendation)

Signal resolution

0.3 µA

Error signal of current

output (adjustable)

High alarm ≥21 mA

Low alarm ≤3.6 mA, last valid measurement

Max. output current

21.5 mA

Starting current

≤10 mA for 5 ms after switching on, then ≤3.6 mA

Damping

0...999 seconds, adjustable

Approvals and certificates

CE

The device fulfils the statutory requirements of the EC directives. The manufacturer

certifies that these requirements have been met by applying the CE marking.

Electromagnetic

compatibility (EMC)

EMC conformity for EN 61326-1 (05/2006)

NAMUR

NE 21 - Electromagnetic compatibility of equipment

NE 43 - Signal level for the failure information of digital transmitters

NE 53 - Compatibility of field devices and display/adjustment components

NE 107 - Self-monitoring and diagnosis of field devices

Classification according to

Pressure Equipment

Directive (PED 97/23/EC)

For gases of fluid group 1 and liquids of fluid group 1, the requirements are fulfilled

according to article 3, paragraph 3 (sound engineering practice).

.book Page 65 Friday, May 2, 2014 12:49 PM