5.4 Extended adjustment

The main menu is divided into five sections:

• Start-up
• Display
• Diagnosis
• Additional adjustments
• Info

5.4.1 Start-up

Adjustment level

1 Sensor mounting correction

Sensor mounting correction

Sensor mounting correction

Sensor mounting correction
In this menu item you compensate the influence of the installation position of the device (offset) on the
measured value.

2 Min.-adjustment

Min.-adjustment

Min.-adjustment

Min.-adjustment
Enter the (pressure) value for the min. level.
At 0% this corresponds to the output signal of 4 mA.

3 Max.-adjustment

Max.-adjustment

Max.-adjustment

Max.-adjustment
Enter the (pressure) value for the max. level.
At 100% this corresponds to the output signal of 20 mA.

Measurement

loop name

Assign a unique device ID. This is useful or even necessary in digital systems and for monitoring large

systems.

Application

In this menu item you select the application: process pressure and level are available. The default

setting is process pressure.

Units

In this menu item the adjustment units, as well as the temperature unit are determined. The selection

of the adjustment unit determines the unit displayed in the "Min-adjustment" and "Max-adjustment"

items. In "Level" mode, it is possible to carry out the adjustment in a height unit (e.g. meters). To do

so, the density of the medium must also be specified.

Adjustment units

Process pressure

Process pressure

Process pressure

Process pressure

mbar, bar, Pa, kPa, MPa, psi, mmH2O,

mmHg, inH2O, inHg and user-defined

Level

Level

Level

Level

Density input required:

mm, cm, m, in and ft

Temperature unit

°

C, °F and K