5.8.1 Error codes
Failure
Check function
Code
Text message
Cause
Action or description
F013
No measurement value available
No valid measured value
available
Overpressure, underpressure,
measuring cell defective
F017
Adjustment span too small
Adjustment not within
specification
Change adjustment according to
the limit values
F025
Error in the linearisation table
Index markers are not
continuously rising, for example
illogical value pairs
Check linearisation table,
Delete table/Create new
F36
No operable sensor software
Failed or interrupted software
update
Repeat software update
Check electronics version
Exchange the electronics or send
device for repair
F40
Error in the electronics
Hardware defect
Exchange the electronics or send
device for repair
F041
Error in the electronics
No connection to sensor
electronics
Check connection to sensor
electronics (with remote version)
F080
General software error
Disconnect operating voltage
briefly
F113
Communication error with the
display and adjustment module,
operating software PACTware
TM
disturbed
EMC interference
Remove EMC influences
F125
Inadmissible electronics
temperature
Temperature of the electronics in
the non-specified section
Check ambient temperature
Isolate electronics
Use device with higher
temperature range
F260
Error in the calibration
Error in the calibration carried
out in the factory
Error in EEPROM
Exchange the electronics
Send device for repair
F261
Error in the configuration
Error during setup
Error when carrying out a reset
Repeat setup,
Repeat reset
F265
Measurement function disturbed
Sensor no longer carries out a
measurement
Carry out a reset
Disconnect operating voltage
briefly
F266
Inadmissible operating voltage
Operating voltage is below the
specified range
Check electrical connection - if
necessary, increase operating
voltage
Code
Text message
Cause
Action or description
C700
Simulation active
A simulation is active
Finish simulation
Automatic end after 60 minutes