Technical data, 4 measurement limits – KROHNE OPTIFLEX 1100C EN User Manual
Page 64
8
TECHNICAL DATA
64
OPTIFLEX 1100 C
www.krohne.com
01/2012 - 4001721801 - HB OPTIFLEX 1100 R01 en
8.4 Measurement limits
Measurement limits (dead zone) in mm and inches
Measurement limits (non-linearity zone) in mm and inches
80 is ε
r
of water; 2.3 is ε
r
of oil
Figure 8-3: Measurement limits
1 Device with a coaxial probe
2 Device with a single cable probe
3 Top dead zone:
Top dead zone:
Top dead zone:
Top dead zone: Top part of the probe where measurement is not possible
4 Top non-linearity zone:
Top non-linearity zone:
Top non-linearity zone:
Top non-linearity zone: Top part of the probe with a lower accuracy of +40 mm/-10 mm / +1.57¨/-0.39¨
5 Bottom non-linearity zone:
Bottom non-linearity zone:
Bottom non-linearity zone:
Bottom non-linearity zone: Bottom part of the probe with a lower accuracy of ±30 mm / ±1.18¨
6 Bottom dead zone:
Bottom dead zone:
Bottom dead zone:
Bottom dead zone: Bottom part of the probe where measurement is not possible
7 Gas (Air)
Gas (Air)
Gas (Air)
Gas (Air)
8 Product
Product
Product
Product
9 L, Probe length
L, Probe length
L, Probe length
L, Probe length
10 Tank Height
Tank Height
Tank Height
Tank Height
11 Minimum distance from the probe to a metallic tank wall:
Minimum distance from the probe to a metallic tank wall:
Minimum distance from the probe to a metallic tank wall:
Minimum distance from the probe to a metallic tank wall: Coaxial probe = 0 mm / 0¨. Single cable probes = 300 mm /
12¨.
Dead zone
ε
r
= 80
ε
r
= 2.3
Top 3
Bottom 6
Top 3
Bottom 6
[mm] [inches] [mm] [inches] [mm] [inches] [mm] [inches]
Coaxial probe
100
3.94
30
1.18
150
5.91
170
6.69
Single cable probes
100
3.94
110
4.33
150
5.91
110
4.33
Non-linearity zone
ε
r
= 80
ε
r
= 2.3
Top 4
Bottom 5
Top 4
Bottom 5
[mm] [inches] [mm] [inches] [mm] [inches] [mm] [inches]
Coaxial probe
250
9.84
30
1.18
250
9.84
30
1.18
Single cable probes
250
9.84
80
3.15
250
9.84
140
5.51