The display

reading does not

change and the

current output

value is incorrect.

2

ERR 200

Reference

pulse lost

Reference amplitude

is less than reference

threshold. This fault

could occur because

the device hardware is

defective.

Contact your supplier to check

that the electronics are still

functioning correctly.

1

ERR 201

Sensor

voltage failure

The device's hardware

is defective.

Contact the supplier.

The display

reading does not

change and the

current output

value is incorrect.

4

ERR 202

Level lost

error

The device cannot find

the product surface.

Measure the level of the contents

in the tank using another method

of measurement. If the tank is

empty (the level is below the end

of the probe), then fill the tank

until product level is in the

measurement range. If the tank is

full (the level is in the blocking

distance), then remove the

contents from the tank until the

level is back in the measuring

range. If the product was lost and

the tank is neither full nor empty,

wait for the device to find the level

again.

The device cannot find

the level return signal

Refer to MEAS. AMP. (Measured

Pulse Amplitude, menu item

2.5.6) and then adjust MEAS.

THRESH. (measurement

threshold, menu item 2.5.7).

The measurement

and the current

output value stops

at the "full" value.

4

ERR 204

Overfill error

The tank is full. The

level signal is lost in

the full area. The

measurement stops at

the blocking distance

value.

Danger of overflow! Remove the

contents from the tank until it is

at a safe level.

The measurement

stops at the last

measured value

and the current

output value is

incorrect.

3

ERR 205

Measurement

old

This error occurs if

the power supply is

not sufficient to

operate the device

correctly for

approximately 30

seconds. After 1

minute the error

message

"Measurement old" is

recorded.

Make sure that the power supply

has a min./max. value of

14...30 VDC when current output

is at 22 mA.

3

ERR 206

No sensor

detected

This error occurs if

there is a problem

with the power supply.

Contact the supplier.

1

Defective wiring.

1

ERR 207

Sensor

memory

failure

The device's hardware

is defective.

Contact the supplier.

1

ERR 208

Oscillator

failure

The device's hardware

is defective.

Contact the supplier.

Problem

Status

marker

shown

Error

code

recorded

Error

Message

Cause

Corrective action