Operation, 2 error handling – KROHNE OPTIFLEX 1100C EN User Manual
Page 54
6
OPERATION
54
OPTIFLEX 1100 C
www.krohne.com
01/2012 - 4001721801 - HB OPTIFLEX 1100 R01 en
6.5.2 Error handling
How to find an error record
• Press [>>>>] and [
] to enter configuration mode from normal mode.
• Press [>>>>], 9 × [
] and [>>>>] to go to menu item 2.10.0 ERROR RECORD.
• Press 2 × [>>>>] to look at the error list. Press [
] or [
] for the selection of an error.
• Press [>>>>] to read more data.
i
The error record gives the number of times the error occurred and the time since the last
error message.
Description of errors and corrective actions
Figure 6-6: Error record data
1 Error code for the error
2 Number of times the error occurred
3 Time since the last error record
INFORMATION!
The time since the error occurred is measured in Days (D), Hours (H), Minutes(') and Seconds (").
It only includes the time when the device is energized. The error is saved in the memory of the
device when it is de-energized. The counter continues when the device is energized again.
Problem
Status
marker
shown
Error
code
recorded
Error
Message
Cause
Corrective action
The measurement
stops at the last
measured value
and the current
output value is
incorrect. The
optional display
screen switches
off automatically.
3
ERR 102
Temperature
out of range
The ambient
temperature is outside
the given range. This
can cause loss or
corruption of data.
De-energize the device until the
ambient temperature is back in
the given range (-40...+80°C /
-40...+176°F).
The measurement
stops at the last
measured value
and the current
output value is
incorrect.
1
ERR 103
Converter
memory
failure
The device's hardware
is defective.
Contact the supplier.
1
ERR 104
Converter
voltage failure
The device's hardware
is defective.
Contact the supplier.