6

OPERATION

40

OPTIFLEX 1100 C

www.krohne.com

01/2012 - 4001721801 - HB OPTIFLEX 1100 R01 en

2.1.3

SEN.FIRM.VER

The sensor firmware version.

Read only.

2.1.4

HMI.FIRM.VER

The HMI (device display screen)

firmware version.

Read only.

2.2.0

TESTS

The device serial number.

2.2.1

SET OUTPUT

This sets the analog output to a

test value [mA] selected from a

list. The output will change to

the selected value after 5

seconds, independent of the

measured value.

3.5, 4, 6, 8, 10, 12, 14,

16, 18, 20 or 22 mA

3.5 mA

2.2.2

DIAGNOSTIC

This starts the hardware test.

Press [>>>>] many times to show:

the time of operation (D1),

temperature of the electronic

converter board (T1), loop

current (I1), load current (I2),

voltage 5.6 V (V1), voltage on

capacitors (V2), voltage 3.3 V

(V3), amplitude reference pulse

(P1), amplitude level pulse (P2),

amplitude probe end pulse (P3),

reset counter (C1). If you press

[>>>>] again, the display goes back

to the menu level.

2.3.0

BASIC PARAM.

The basic parameters for

configuration of the device.

2.3.1

TANK HEIGHT

The distance from the thread

stop down to the tank bottom.

Probe length

2.3.2

TIME CONST.

Increasing the time constant

will smoothen the integrated

readings, decreasing will

roughen the readings.

min-max:

0 to 100 seconds

5 seconds

2.3.3

PROBE LENGTH Probe length is the distance

from the thread stop of the

device down to the bottom end

of the probe (including

counterweight for cable

versions). If probe length has

been modified, enter the new

value here. For more data,

refer to

Quick Setup

on page 44.

min: 2.3.4 BLOC.

DIST. + 3.1.1

COUNTERWEIGHT

max.: 20 m / 65.6 ft

1

2.3.4

BLOC. DIST.

Blocking distance. The non-

measuring range at the top of

the probe. It depends on the

probe type and the installation.

Make sure that the product

does not go above the blocking

distance. There is a risk that

the product will overflow

and/or cover the device.

min: 0 m / 0 ft

max: 2.3.3 PROBE

LENGTH

50 mm / 2¨

2.3.5

TAG NAME

The device has a code (tag

name) to identify it.

2.4.0

OUTPUT I

Use this sub-menu to give

measurement values to the

current output.

Menu
No.

Function

Function description

Selection list

Default