Installation – KROHNE OPTIFLEX 1100C EN User Manual

Page 18

background image

3

INSTALLATION

18

OPTIFLEX 1100 C

www.krohne.com

01/2012 - 4001721801 - HB OPTIFLEX 1100 R01 en

• Measure the probe (with all the segments attached).
• Mark the probe length necessary for your application on the tube.
• Remove the last segment of tube.

Adjust the probe length to the tank height

Figure 3-12: How to assemble the segmented coaxial probe: part 2

a = Delivered probe length (with all segments attached, but without the probe end cap)
b = Length of tube needed = total probe length, L - 15 mm / 0.6¨
1 Coaxial probe: Last segment of the tube
2 Do not attach the lock screw to the last segment of tube! Make sure that the coaxial probe is the correct length for

the tank.

3 Scriber mark - to cut the tube to the correct length

CAUTION!
If it is necessary to decrease the length of the probe, do not attach the lock screw on the last
segment of tube.
Make sure the probe is the correct length for the tank. Add the probe end cap length (15 mm /
0.6

¨

) to get the total length of the probe.