6.5 Status and error messages

6.5.1 Status markers

If the device senses a change in device status, the display screen will show 1 or more status
markers.

Status markers

For data on errors, refer to

Error handling

on page 54.

Figure 6-5: Status markers

1 Status marker line
2 When the status marker is on, a number is shown

Code

Name

Description

1

Hardware problem

The device's hardware is defective. Contact the supplier.

2

No reference pulse

Reference amplitude is less than reference threshold. This

fault could occur because the device hardware is defective.

For more data, refer to

Error handling

on page 54.

3

Low voltage /

Measurement old

This error occurs if the power supply is not sufficient to

operate the device correctly for 1 minute. After 1 minute the

error message "Measurement old" is recorded. The output

current changes to an error value after a time given in menu

item 2.4.5 ERROR DELAY. For more data, refer to

Error

handling

on page 54.

There is a problem with the connection between the sensor

and the device converter. Contact the supplier.

4

Level lost

The device cannot find the product surface or the end of the

probe (when ε

r

<1.6). The measurement stops at the last

measured value.