SUMMIT 8800

CALIBRATION

08

Figure 95 Analog output calibration screen

Analog Output Calibration Procedure

NOTE: Output accuracy is specified as 0.15% FSD or an absolute error of 30 uA.

• Connect a reference precision 100 Ohm resistance (0.05% accuracy or better) to the target

analog output.

• Connect a reference calibrated digital meter with a 0-2V range (0.05% accuracy or better)

across the resistor.

• Click the SET TO 4mA button on the START CALIBRATE page.

• For below two adjustments allow for input value stabilisation (e.g. 30 seconds) and assume

that stabilisation is reached when output values have oscillated 5 times around the assumed

final value.

• Use the up and down arrow buttons to raise and lower the output until 0.4000V is indicated on

the meter.

• Click the SET TO 20mA button on the START CALIBRATE page.

• Use the up and down arrow buttons to raise and lower the output until 2.0000V is indicated on

the meter.

• Repeat above two steps 3 times over. Click the WRITE button to store the results.

• Calibrate the output by simulating 4mA. Does the output value field read 0.4000V? Simulate

100%.

• Does the output value field read 2.0000V?

• Click the STOP CALIBRATE button to exit the analog output calibration menu.

8 .2 .2 Digital Output

The Digital Outputs do not require any customer calibration.