SUMMIT 8800

CALIBRATION

08

• Apply the current that corresponds with the minimum process value to the unit. Note that

only 0% (4mA) and 100% (20mA) values will be accepted. It is not possible to calibrate a range

with e.g. 40% and 75% as minimum and maximum calibration values. Set the field instrument

in simulation mode and simulate 0% (4mA). For below two adjustments allow for input value

stabilisation (e.g. 30 seconds) and assume that stabilisation is reached when input values

have oscillated 5 times around the assumed final value. In above example: type in -50C and

click the upper SET button. Simulate 100% (20mA) and click the lower SET button. Repeat

above two steps 3 times over. Click the WRITE button to store the results.

• Calibrate the input by simulating 0%. Does the uncalibrated value field read -50 degrees Cel-

sius? Simulate 100%.

• Does the uncalibrated value field read 150 degrees Celsius?

• Click the END CALIBRATION AND CLOSE button to leave the analog input calibration menu.

Analog Input Calibration Procedure

• Apply the Lower Calibration input to the unit (e.g. set the 4-20mA Input to 4.8mA i.e. 5% - 5

bar)

• Type in 5.0 to the Wanted box for Point 1 above.

• Press the Set button for Point 1

• Wait for stabilization.

• Apply the Upper Calibration input to the unit (e.g. set the 4-20mA Input to 19.2 mA i.e. 95% -

95 bar)

• Type in 95.0 to the Wanted box for Point 2 above.

• Wait for Stabilization.

• Press the Set button for Point 2.

It will be necessary to repeat the above steps a number of time until the predicted value is stable

and correct for both the lower and upper calibration points.

Once the Calibration is correct and stable for both points press the Write button to commit the

calibration to memory.

Select the Close button to exit the Analog Input Calibration menu.

8 .1 .4 Digital Input

The Digital Inputs do not require any customer calibration.

8 .2 Output Calibration

8 .2 .1 Analog Output

The factory calibrated analog output signals normally do not require customer adjustment. It is

recommended that the SUMMIT 8800 outputs be calibrated and adjusted if necessary.

Main Controls:
Start calibrate

starts the calibration process

Write

write the new calibration data to the connected flow computer

Discard

discards any calibration changes made.

Default

reverts to default factory calibration.

Stop calibrate

stops the calibration process.