4 .2 .5 Back Up Battery

The SUMMIT 8800 contains a backup battery on the auxiliary card, this battery keeps the real

time clock and internal totals when DC power to the unit is disconnected. Replacement of this

battery should only be carried out by qualified personnel.

If the SUMMIT 8800 is kept continuously powered, the back-up battery estimated life is 10 years.

If the SUMMIT 8800 is powered down for periods exceeding 30 days, the battery should be re-

moved and stored separately. The battery needs to be replaced when its service life exceeds 10

years or when it has been left in an un-powered unit for a time period of greater than 2 months.

A “BAD” battery condition is indicated via the Windows software.

When a “BAD” battery condition is indicated via the Windows software or on the front panel: