DECIMAL

Decimal (Point)

R

The number of digits to the right of the decimal point which

should be used by an interface device in displaying the

specified parameter.

2

SELECTOR_DIAG

SELECTOR_DIAG

SELECTOR_DIAG

SELECTOR_DIAG

selector diagnosis

R/W

Selector for diagnosis values.

-

SELECT_DIAG1

diagnostics 1

R/W

Options for diagnosis 1 (Fct. C1.16.1)

off

Settings:

off / velocity of sound 1 / gain 1 / SNR 1

SELECT_DIAG2

diagnostics 2

R/W

Options for diagnosis 2 (Fct. C1.16.2)

off

Settings:

off / velocity of sound 2 / gain 2 / SNR 2

SELECT_DIAG3

diagnostics 3

R/W

Options for diagnosis 3 (Fct. C1.16.3)

off

Settings:

off / velocity of sound 3 / gain 3 / SNR 3

TRANSDUCER_

TRANSDUCER_

TRANSDUCER_

TRANSDUCER_

DIRECTORY

DIRECTORY

DIRECTORY

DIRECTORY

Transducer

Directory Entry

A directory that specifies the number and starting indices of

the data collections in the transducer block.

0

TRANSDUCER_

TRANSDUCER_

TRANSDUCER_

TRANSDUCER_

TYPE

TYPE

TYPE

TYPE

Transducer Type

R

Identifies the transducer that follows.

Other

XD_ERROR

XD_ERROR

XD_ERROR

XD_ERROR

Transducer Error

One of the transducer error codes defined in the FF

Transducer Specifications in section "4.7 Block Alarm

Subcodes".

0

Settings:

16: Unspecified error

17: General error

18: Calibration error

19: Configuration error

20: Electronics Failure

21: Mechanical Failure

22: I/O Failure

23: Data Integrity Error

24: Software Error

25: Algorithm Error