beautypg.com

In-Situ Aqua TROLL 100 and 200 Operators Manual User Manual

Page 86

background image

Page 86

Aqua

TROLL Operator’s Manual

Section 10: Troubleshooting

• For a log with a scheduled start, select None as the stop

condition, or select a stop time that is closer to the start

time.

TROUBLESHOOTING PARAMETER CONFIGURATION
Problem:
Cannot configure level or other parameters using

the Configure button on the Sensors tab. The Sensor setup

screen is shown, but the Configure... button is dim.

Probable Cause 1: The Aqua TROLL is actively polling

(continually updating real-time readings) in the Home tab.

Suggested Remedy: Return to the Home tab and stop real-time

readings by clicking

.

Probable Cause 2: The Aqua TROLL has an active log—a log that

is Ready, Pending, Running, or Suspended as shown in the

Status column of the Logging Tab. Only one active log can

reside in the device at a time.

Suggested Remedy: Stop or delete the log if possible.

Alternatively, configure parameters after the log is

completed.

TROUBLESHOOTING CONDUCTIVITY CALIBRATION
Problem:
The calculated cell constant (Kcell) is lower than the

recommended range of 0.98 – 1.02.

Probable Cause: The electrical resistance of the calibration

solution was lower than expected for the nominal value.

Suggested Remedy: Check the following and then repeat the

calibration:

• Visually inspect the sensor for the presence of fouling

(most likely conductive material, possibly a bio-film);

clean as described in Section 9 of this manual.

• Ensure the calibration solution is not contaminated,

mislabeled, or partially evaporated.

This manual is related to the following products: