Internal Power
The Aqua TROLL Instrument operates on 3.6 VDC, supplied by a
sealed, non-replaceable AA lithium battery. Battery life depends
on sampling speed. The battery typically lasts for 5 years or
200,000 readings, whichever occurs first. One reading is defined
as date, time, and all available parameters polled or
logged from the device.
External Power
External Battery Pack
The sealed, submersible TROLL Battery Pack (lithium)
supplies 14.4 V. When this power source is connected,
the Aqua TROLL Instrument will use the external battery
source first and switch to the internal batteries when
external battery power is depleted. Typical external
battery life when logging all available parameters at
2-minute intervals is approximately 2 years.
AC Adapter
In-Situ’s AC adapter provides 24 VDC, 0.75 A, AC input 100-
250 V, includes North American power cord. The Programming
Cable includes an external power input for connection to this
adapter.
Accessory
Catalog No.
External Battery Pack ......................................................................................0051450
AC Adapter 24V .................................................................................................0052440
TIP: Win-Situ
5 can display
the approximate
percentage of internal battery
life remaining when the Aqua
TROLL Instrument is connected
to a computer.
TIP: When an Aqua
TROLL Instrument is
used as an Analog
(4-20 mA), SDI-12, or Modbus
device, power is supplied by
the data logger or controller
to which the Aqua TROLL
Instrument is wired.
Use only In-Situ’s AC
adapter. Damage
to the Aqua TROLL
Instrument caused by the use
of third-party converters is not