7 Field Installation
Positioning the Vented Aqua TROLL 200
Instrument
For unattended deployment, lower the Aqua TROLL 200
Instrument gently to approximately the desired depth. Position
the instrument below the lowest anticipated water level, but
not so low that its range might be exceeded at the highest
anticipated level. Refer to the tables below for usable depth.
Range
Usable Depth
psig
kPa
Meters Feet
5 34.5
3.5 11.5
15 103.4
11
35
30 206.8
21
69
100
689.5
70
231
300
2068
210
692
500
3447
351
1153
Vented Aqua TROLL Instrument
Non-Vented Aqua TROLL Instrument
Range Effective Range* Usable Depth
psia
psia
kPa Meters
Feet
30
15.5 106.9
10.9
35.8
100
85.5 589.5
60.0
197
300
285.5 1968
200.7
658.7
500
485.5 3347
341.3
1120
* At sea level (14.5 psi atmospheric pressure).
Do not let the
instrument fall
freely into the
water because it can damage
the pressure sensor.