beautypg.com

7 field installation, Positioning the vented aqua troll 200 instrument – In-Situ Aqua TROLL 100 and 200 Operators Manual User Manual

Page 67

background image

Page 67

Aqua

TROLL Operator’s Manual

7 Field Installation

Positioning the Vented Aqua TROLL 200

Instrument
For unattended deployment, lower the Aqua TROLL 200

Instrument gently to approximately the desired depth. Position

the instrument below the lowest anticipated water level, but

not so low that its range might be exceeded at the highest

anticipated level. Refer to the tables below for usable depth.

Range

Usable Depth

psig

kPa

Meters Feet

5 34.5

3.5 11.5

15 103.4

11

35

30 206.8

21

69

100

689.5

70

231

300

2068

210

692

500

3447

351

1153

Vented Aqua TROLL Instrument

Non-Vented Aqua TROLL Instrument

Range Effective Range* Usable Depth
psia

psia

kPa Meters

Feet

30

15.5 106.9

10.9

35.8

100

85.5 589.5

60.0

197

300

285.5 1968

200.7

658.7

500

485.5 3347

341.3

1120

* At sea level (14.5 psi atmospheric pressure).

Do not let the
instrument fall
freely into the

water because it can damage
the pressure sensor.

This manual is related to the following products: