TROUBLESHOOTING CONNECTIONS
Problem:
Win-Situ 5 Software cannot connect to the Aqua

TROLL instrument.

Probable Cause: The wrong COM port is selected,

communication settings are incompatible, batteries are low,

or cable connections are loose or dirty.

Suggested Remedy: Check the following:

• All cable connections are tight, connectors are clean

and dry.

• The cable is securely attached to the instrument.

• The correct COM port is selected (select Preferences

menu > Comm Settings to check this).

• The communication settings in Win-Situ and in the

Aqua TROLL match. To reset the device communication

settings to the serial defaults, click “Reset all Devices” in

the Comm Settings dialog (Preferences menu > Comm

Settings)

• The internal battery has voltage remaining or external

power is supplied.