Aqua
TROLL Operator’s Manual
Check Instrument Depth
At this point, if convenient, you can connect the Aqua
TROLL 200 to a PC, launch the software, and take a
reading. If the instrument is at the desired depth, secure it
in position as suggested below, or reposition as necessary.
If you set the software to “Remind me later” to set a Level
Reference, enter the level reference after installation when
prompted.
Secure the Cable
The RuggedCable includes a Kellems® grip near the
surface end. You can slide it along the cable to the desired
position by compressing it. When you pull on it, the grip
tightens and stops sliding. You may need to pull on both
ends of the Kellems grip to properly tighten it and keep it
from slipping.
Use the loop of the Kellems grip to anchor the cable to a
convenient stationary object. It works well with In-Situ’s
well dock installation ring. Simply insert the loop into the
locking clip on the well dock, and position the assembly
on the top of a well.
Installation Tips
4
Never let a probe fall freely down a well. The resulting
shock wave when it hits the water surface can damage
the strain gauge.
4
Check the level of water above the probe, then move
it and read again to be sure that the probe presents a
reasonable reading and shows change. It might not
be located where you think it is. For example, it could
Section 7: Field Installation
Kellems grip