Section 4: Using Win-Situ

Check the Conductivity Calibration
The conductivity sensor in the Aqua TROLL has been calibrated

during manufacturing to produce a linear response across

the operating range.

We recommend you check the specific conductivity

reading in the solution shipped with your instrument. If

the device is reading accurately, there is no need to field-

calibrate the instrument unless SOPs require it.

To perform a field calibration you will need:

4

The In-Situ Cal Cup, or other suitable container that

allows for complete immersion of the conductivity

sensor, including the temperature “button,” which

can be seen on one side of the probe. The Cal Cup is

recommended for the first calibration.

4

The calibration standard solution

supplied, or other solution of known

specific conductivity in the range 100 to

60,000 µS/cm.

The complete calibration procedure takes

only a few minutes and may be done in a

field or office/lab setting with a software connection in Win-Situ

5 or Win-Situ Mobile Software. A detailed description of the

conductivity calibration procedure, including:

• Preparation of the Aqua TROLL

• Software input

• Calculated output

• Hints for successful calibration

is in

Section 5

of this manual.

For best results we
recommend you
read and follow

the conductivity calibration
procedure in Section 5 of this
Operator’s Manual.

Temperature

Conductivity

Pressure