Batteries
Internal batteries in the Aqua TROLL Instrument are not user-

replaceable. The approximate percentage remaining is displayed

on the Win-Situ 5 Dashboard when the instrument is connected.

Cleaning
Probe and Front End
Clean the Aqua TROLL with water and a soft brush or plastic

scouring pad, or soak overnight in a mild acidic solution, such as

household vinegar.

If the pressure ports of the Aqua TROLL 200 in the front end are

clogged with silt or mud, try the following:
• Agitate the instrument vigorously in a bucket of clean water.

• Apply a gentle stream of water from a wash bottle.

• In severe cases, remove the nose cone and clean out the

holes with a soft brush or pipe cleaner.

To avoid damage to the pressure sensor diaphragm, do not insert

any object into the sensor opening or attempt to dig out dirt or

other materials.

If contamination cannot be removed using the

recommendations above, please contact In-Situ Inc. for cleaning.

Section 9: Care & Maintenance

When the nose
cone is removed
from the Aqua

TROLL 200, the sensitive pressure
sensor diaphragm is completely
exposed. Do not touch this area
with any object. Replace the
nose cone as soon as possible.

Nose cone removed - do

not touch the exposed

diaphragm

Aqua TROLL 200 front

end with nose cone

in place

Damage caused by digging or scraping in the pressure sensor opening to remove
silt, mud, etc. is not covered by the warranty.

TIP: If batteries
are completely
exhausted,

remember that external power
and battery pack options are
available.