Rear panel – GW Instek PSH-A Series 360W User Manual User Manual
Page 9
PSH
Overview
17
5 Vset/
Iset/
Enter key
Switches between voltage setting mode and
current setting mode, or confirm the entered
value in the menu mode (see page44).
Vset (edit Voltage)
OCP
SET
2.58V
2.01A
Iset (edit Current)
OCP
SET
2.58V
2.01A
6 F/C
(Fine/
Coarse) key
Switches the editing location and resolution:
before (coarse) or after (fine) the decimal point.
Fine
(after
decimal)
OCP
SET
2.58V
2.01A
Coarse (before decimal)
OCP
SET
2.58V
2.01A
7 Menu
key
Enters into the menu mode. For details, see
page44.
Default
mode
OCP OUT
CV
2.58V
2.01A
Menu mode (OVP setting)
Set OVP
* 21.10V
8 Local
key
Switches from remote control mode to local
operation mode (page56), OR releases OVP/OCP
error messages and go back to normal operation
(page37), OR enters the calibration mode when
pressed for more than 5 seconds (page61).
Error message example
OCP Error! Press
“Local” to reset
Calibration mode
Calibration
Voltage
PSH
User
Manual
18
Rear Panel
1 Ground
Terminal
Connect the output line shield (page25) and
the remote sensing line shield (page32).
2 Output
Terminal
Connect DUT (Device Under Test). For details,
see page24.
3 Sense
Terminal
Connect the feedback line to compensate for
cable loss. For details, see page24 (theory),
page32 (cable connection).
4 RS232/
GPIB
Terminal
Connect the remote control line. For remote
control details, see page52. GPIB requires an
optional module. For installation details, see
the service manual.
Note: Only one interface module (RS232 or
GPIB) can be installed at a time.
5 AC
input
Terminal
Connect the AC power input cable. For
details, see page22.