Run calibration when the output exceeds the
specification, or when the functionality check
(page34) fails. To get the optimal result, make
sure PSH is powered for at least 30 minutes

before the calibration.

Setup

Calibration Preparation ............................. 62

Entering calibration mode ......................... 63

Calibration

Output Voltage calibration......................... 64

Output Current calibration......................... 66

OVP Calibration ....................................... 68

PSH

User

Manual

62

Calibration Preparation

Calibration
condition

At least 30 minutes of warm-up time

Temperature: 23 ± 5 °C

Relative Humidity: ≤ 80%

Calibration
items

Minimum/ Maximum Output Voltage

Minimum/ Maximum Output Current

OVP functionality

Digital

Multimeter

DC Voltage Accuracy: <±0.1%

Recommended model:
GDM-8245, GDM-8246

Current Shunt

Current range: >100A

Accuracy: <±0.1%

PSH – Current
Shunt cable

Voltage rating: >70V

Current rating: >100A

Calibration
Equipment

Multimeter –

Current Shunt
cable