Calibration

65

5. Adjust the display value to the

Multimeter reading.
Wheel knob

changes the value.

F/C key

F/C

moves the cursor

before and after the decimal point.

Multimeter reading:
20.03V

Calibration

Voltmax 20.03V

6. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

to confirm the setting.

The display goes back to the
default calibration menu.

Calibration

Voltage

7. Save the result and exit the

calibration mode (Save), OR

Exit the calibration mode without
saving the result (Exit).

Calibration

Save

Calibration

Exit

PSH

User

Manual

66

Output Current calibration

Connection

Digital Multimeter

PSH Series

V

A

— +

Current

Shunt

Calibration step

1. Enter the Current calibration

menu. See page63 for details.

Calibration

Current

2. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

and enter Maximum

Current calibration mode.

Calibration
CurrMax 18.00A

3. Adjust the display value to the

Multimeter / Current Shunt
reading.
Wheel knob

changes the value.

F/C key

F/C

moves the cursor

before and after the decimal point.

Multimeter reading:18.02A

Calibration
CurrMax 18.02A

4. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

to confirm the setting.

The display enters Minimum
Current calibration mode.

Calibration

CurrMin 0.00A