Calibration

63

Entering calibration mode

1. Press the Local key for 5 seconds.

The password entry menu

appears.

Please enter the

Password: 0

2. Enter the model number as the

password.
Wheel knob

changes the digit.

F/C key

F/C

moves the cursor to

the next digit.

(PSH-2018A)

Please enter the

Password: 2018

3. Press the Vset/Iset/Enter key

V

SET

/ I

SET

ENTER

as confirmation.

4. The calibration menu appears.

Wheel knob

selects the item.

Vset/Iset/Enter key

V

SET

/ I

SET

ENTER

confirms the selection.

Minimum/Maximum output

voltage calibration

Minimum/Maximum output
current calibration

Over Voltage Protection calibration

Save the calibration result and exit
the calibration menu

Exit calibration menu without

saving the calibration result

Calibration

Voltage

Calibration

Current

Calibration

OVP

Calibration

Save

Calibration

Exit

PSH

User

Manual

64

Output Voltage calibration

Connection

Digital Multimeter

PSH Series

V

A

+

Calibration step

1. Enter the Voltage calibration

menu. See page63 for details.

Calibration

Voltage

2. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

and enter Minimum

Voltage calibration mode.

Calibration

Voltmin

0.00V

3. Adjust the display value to the

Multimeter reading.
Wheel knob

changes the value.

F/C key

F/C

moves the cursor

before and after the decimal point.

Multimeter reading: 0.02V

Calibration

Voltmin

0.02V

4. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

to confirm the setting.

The display enters Maximum

Voltage calibration mode.

Calibration

Voltmax 20.00V