Setup

35

Output Voltage & OVP Check

Connection

Digital Multimeter

PSH Series

V

A

+

Checking step

1. Power On PSH and connect the Multimeter Voltage

measurement terminal.

2. Set Output Voltage and Output Current to the rating value.

Vset/Iset/Enter key

V

SET

/ I

SET

ENTER

switches the cursor between

Voltage and Current.

F/C key

F/C

switches the cursor

between before and after the
decimal point.

Wheel knob

changes the value.

PSH-2018A (20V, 18A)

SET 20.00V

18.00A

3. Press the Menu key once and set the OVP value to rating

voltage + 100mV.

Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

to confirm the OVP

setting.

Set OVP
20.10V

PSH

User

Manual

36

4. Make sure the Multimeter shows no value (No output).

5. Press the Output key

OUTPUT

and turn On the output.

The display changes into CV

(Constant Voltage) mode and
shows the OUT sign on the top

right corner.

OUT

CV

20.00V

18.00A

6. Make sure the Multimeter reading and the PSH display show

approximately the same Voltage.

7. Increase the PSH output voltage beyond the OVP value.

8. Make sure the display shows an error message.

Press the Local key

LOCAL

to get

back to the normal display.

OVP Error! Press
“Local” to reset

9. Make sure the output is now turned Off in the display and the

Multimeter shows no value (No output).