Ovp calibration – GW Instek PSH-A Series 360W User Manual User Manual

Page 34

background image

Calibration

67

5. Adjust the display value to the

Multimeter / Current Shunt
reading.
Wheel knob

changes the value.

F/C key

F/C

moves the cursor

before and after the decimal point.

Multimeter reading: 0.02A

Calibration

CurrMin 0.02A

6. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

to confirm the setting.

The display goes back to the
default calibration menu.

Calibration

Current

7. Save the result and exit the

calibration mode (Save), OR
Exit the calibration mode without

saving the result (Exit).

Calibration

Save

Calibration

Exit

PSH

User

Manual

68

OVP Calibration

Connection

No connection required

Calibration step

1. Enter the Voltage calibration

menu. See page63 for details.

Calibration

OVP

2. Press the Vset/Iset/Enter

key

V

SET

/ I

SET

ENTER

. PSH calibrates OVP

automatically.

Calibration

OVP

100%

3. The display goes back to the

default calibration menu.

Calibration

OVP

4. Save the result and exit the

calibration mode (Save), OR
Exit the calibration mode without

saving the result (Exit).

Calibration

Save

Calibration

Exit