beautypg.com

Remote sensing wire assembly – GW Instek PSH-A Series 360W User Manual User Manual

Page 16

background image

Setup

31

Load Wire Assembly (cont.)

4. Terminal
cover assembly

Screw the output terminal cover to the rear
panel.

5. Wire shield
connection
(recommended)

Connect the load wire shield to the rear panel

ground connector.

PSH

User

Manual

32

Remote Sensing Wire Assembly

Select the appropriate wire according to the guideline on page25.

Local sensing

The sense terminal is connected to the output
monitor terminal with bare wires.

Remote Sensing

1. Take off the wire jumpers.

2. Screw wires to the sense (S) side.

CAUTION

Do not screw remote sensing wires to the
monitor (M) side.

M+, M− are for output monitoring only. NEVER
screw load wires to the remote sensing
terminal.