beautypg.com

GW Instek PSH-A Series 360W User Manual User Manual

Page 28

background image

Remote

Operation

55

GPIB (Optional)

Refer to the service manual for GPIB module installation.
Note: The RS232 module has to be replaced with the GPIB

module: they cannot be used together.

Address setting

1. Press the Menu key

MENU

repeatedly until the

Interface menu appears.

OCP OUT

CV

2.58V

2.01A

Interface GPIB
Address 8

2. Use the wheel knob

to select the address.

GPIB address: 1 ~ 30

3. Press the Iset/Vset/Enter key

V

SET

/ I

SET

ENTER

to

confirm the setting.

4. Connect the GPIB cable between your PC and

PSH rear panel. PSH switches to remote
control mode as soon as the connection is
established.

GPIB
constraints

Address range: 1 ~ 30

Altogether less than 15 devices, less than 20m

total cable length, maximum 2m between each
device

Unique address assigned for each device

At least 2/3 of the GPIB devices turned On

No loop or parallel structure

PSH

User

Manual

56

Interface functionality check

Remote mode
display

In the remote control mode, the display shows
“RMT” on the top left corner. The panel operation
is disabled.

RMT

OCP

CV

0.05V

1.05A

To enable panel operation, press the Local

key

LOCAL

. This also disconnects the remote

control.

Functionality
check

Run this query command via the terminal.

*idn?

This should return the Manufacturer, Model
number, Serial number, and Firmware version in
the following format.

GW.Inc,PSH-2018A,12345678,FW1.00