Yokogawa µR20000 User Manual
Page 42
4-11
SM 04P02B01-01E
Adjustments
4
Adjusting the Full Position
9.
Select P_Adj = Full, then press the
key.
The adjustment value input screen in displayed.
P_Adj=Full
10.
Adjust the numerical value so that the pen recorder position is in alignment from
the 0% position of the z-fold chart paper to the 180 mm position.
Full=4
-30/30(B:1000)
11.
When the adjustment is complete, press the
key.
You are returned to the Hysteresis/Zero/Full selection screen.
Saving Adjusted Values
12.
When all adjustments are finished, select Mode = End, then press the
key.
Mode=End
13.
With End = Store displayed, press the
key.
The adjusted value is saved.
End=Store
Concluding Adjustments
14.
Select Menu = End, then press the
key. You are returned to the normal
operating mode.
Menu=End
4.3 Adjustments to the Dot Model
* Select using the key. To enter numerical values, move the cursor to the entry point using the
key. Select the value using the key. You can hold down the SHIFT key while pressing
key or key to reverse the key’s action.