Adjustments

4

Procedure

The following explains the procedure when working with channel 1.
1.

Press the

MENU

and

CH UP

keys while turning ON the power. Menu = A/D

Adjust is displayed.

2.

With Menu = A/D Adjust is displayed, press the

key.

Menu=A/D Adjust

3.

Input the password (see page 3-5).

Password=

4.

Select Mode = Cal/Exec, then press the

key.

Mode=Cal/Exec

5.

Select A/D No. = 1, then press the

key.

*

For channel 3, select A/D No. = 2.

A/D No.=1

6.

Select Range = 20 mVZ.

Range=20mVZ

7.

Enter the zero value for the 20 mV range, then press the

key.

After measurement, the adjusted value is displayed.

Range=20mVZ
Zero:-200

8.

Press the

key.

You are returned to the A/D No. = screen.

9.

Leave the A/D No. unchanged, then press the

key.

A/D No.=1

10.

Select Range = 20 mVF, then press the

key.

Range=20mVF

11.

Enter the Full value for the 20 mV range on channel 1, then press the

key.

After measurement, the adjusted value is displayed.

Range=20mVF
Full:32768

12.

Repeat steps 5 to 11 to adjust the following ranges: 60 mV, 200 mV, 1 V, 2 V, 6
V, 20 V, Pt100, Cu10 (with the /N1 option).

Saving Adjusted Values
13.

When all adjustments are finished, select Mode = End, then press the

key.

Mode=End

14.

Select A/D No. = 1, then press the

key.

*

You can select an A/D No. of 1, 2, or 1+2. The adjusted value is saved on the AD

converter of the selected number.

A/D No.=1

4.2 Adjustments to the Pen Model

* Select using the key. To enter numerical values, move the cursor to the entry point using the
key. Select the value using the key. You can hold down the SHIFT key while pressing
key or key to reverse the key’s action.