Yokogawa µR20000 User Manual
Page 39
4-8
SM 04P02B01-01E
● Procedure
1.
Press the
MENU
and
CH UP
keys while turning ON the power. Menu = A/D
Adjust is displayed.
2.
With Menu = A/D Adjust is displayed, press the
key.
Menu=A/D Adjust
3.
Input the password (see page 3-6).
Password=
4.
Select Mode = Cal/Exec, then press the
key.
Mode=Cal/Exec
5.
Select A/D No. = 1, then press the
key.
A/D No.=1
6.
Select Range = 20 mV.
Range=20mV
7.
Enter the zero value for the 20 mV range on channel 1 and the Full value on
channel 2, then press the
key.
After measurement, the adjusted value is displayed.
Range=20mV
Z:-585 F:32769
8.
Press the
key.
You are returned to the A/D No. = screen.
9.
Repeat steps 5 to 8 to adjust the following ranges: 60 mV, 200 mV, 1 V, 2 V, 6
V, 20 V, Pt100, Cu10 (with the /N1 option).
Saving Adjusted Values
10.
When all adjustments are finished, select Mode = End, then press the
key.
Mode=End
11.
Select A/D No. = 1, then press the
key.
A/D No.=1
12.
With End = Store displayed, press the
key.
The adjusted value is saved to the A/D converter.
End=Store
Concluding Adjustments
13.
Select Menu = End, then press the
key. You are returned to the operation
mode.
Menu=End