beautypg.com

Channel-to-channel error, Burnout detection, Channel-to-channel error -11 – Yokogawa µR20000 User Manual

Page 24: Burnout detection -11

background image

3-11

SM 04P02B01-01E

Testing

3

Channel-to-Channel Error

Procedures

Pen Model
Assign inputs as shown in the table below in parallel to channels 1 and 2, and check
whether the measured values on both channels meets the criteria. Perform the same
test using channels 3 and 4.
Criteria

Range

Input

Criteria

20 mV

Short +/A and −/B

0.00 ±0.01 mV

Pt100

103.9 Ω

10 ±0.2°C, between channel dispersion within 0.2°C

Dot Model
Assign inputs as shown in the above table in parallel to all channels, and check
whether the measured values on each channel meet the test criteria in the table
above.

*

The b terminal of the resistance temperature detector input is connected internally.
However, when option /N2 is specified, the b terminal is isolated at each channel.

Burnout Detection

Procedures

The test is performed on a representative channel

*

from each A/D converter. Set up

the burnout detection function on the recorder in advance.

* On a four-pen model for example, perform the test on channel 1 and channel 3. On a dot

model, check an arbitrary channel.

Check the action in the following cases.
• Connect 2 kΩ between the channel’s +/A terminal and −/B terminal. Set that

channel’s thermocouple type to K and check that a value close to room
temperature is displayed.

• Connect 82 kΩ and 0.1 µF in parallel between the channel’s +/A terminal and −/B

terminals. Set that channel’s thermocouple type to K and check that the displayed
value is “B. out.”

3.4 Test Procedures

This manual is related to the following products: