4 test procedures, Insulation resistance, Protective grounding – Yokogawa µR20000 User Manual
Page 21: Test procedures -8 insulation resistance -8, Protective grounding -8
3-8
SM 04P02B01-01E
3.4
Test Procedures
Insulation Resistance
● Procedures
Perform a measurement with an insulation tester and check whether the results meet
the criteria. Perform the measurement with the power switch turned ON.
Criteria
Terminals Measured
Criteria
Condition
Between the power and earth terminals
100 MΩ or more
Short the L and N terminals.
Between the input and earth terminals
100 MΩ or more
Short all input terminals.
Between input terminals
100 MΩ or more
Short the odd and even channels of
the +/A, −/B, and b* terminals, and
measure between odd and even
channels.
Between the relay contact output
100 MΩ or more
With the /A1, /A2, /A3, /A4, /A5, or
terminals and earth terminals
/F1 option. Short all relay contact
output terminals.
Between remote control input
100 MΩ or more
With the /R1 option.
C terminal and earth terminals
Between the RS-422A/485 SG
100 MΩ or more
With the /C3 option.
terminals and earth terminals
Between the Ethernet terminal
100 MΩ or more
With the /C7 option. Short all
and earth terminals.
input/output terminals
*
With the dot model without the /N2 option, exclude the b terminal. The b terminals are
connected to all channels.
Protective Grounding
● Procedures
Perform a resistance measurement with a DMM (Digital Multi Meter), and check
whether the results meet the criterion.
Criterion
Measured Location
Criterion
Between earth terminal and power supply heat sink
0.2 Ω or less
Heat sink
Protective ground terminal
L
N