2 adjustments to the pen model, Adjusting the a/d converters – Yokogawa µR20000 User Manual
Page 33
4-2
SM 04P02B01-01E
4.2
Adjustments to the Pen Model
Adjusting the A/D Converters
● Overview
Enter the value determined for each range, perform measurement, then write the
adjusted values to the AD converter.
● Instruments Used
Instrument
Specifications
DC voltage generator
Accuracy: ±0.005% of setting + 1 µV or better
Resistors
10 Ω, 50 Ω, 100 Ω, 150 Ω, 300 Ω, Accuracy: 0.01% or better
● Channel
Adjustment of channels 1 and 2 is performed using channel 1, and adjustment of
channels 3 and 4 is performed using channel 3.
● Procedure
Connection
UUA
When DC range
DC Voltage
standard
DECADE
RESISTANCE
BOX
Channel 1 or
Channel 3
H
L
UUA
UUA: Unit Under Adjustment
When RTD or Cu range
− +
Channel 1 or
Channel 3
b B A
Input Value
Input Type
Range
Zero
Full
Description
DC voltage
20 mV
0 mV
20 mV
-
60 mV
0 mV
60 mV
-
200 mV
0 mV
200 mV
-
1 V
0 mV
1 V
-
2 V
0 mV
2 V
-
6 V
0 mV
6 V
-
20 V
0 mV
20 V
-
50 V
No adjustment needed
-
RTD
Pt100
100 Ω
300 Ω
-
Cu10
10 Ω
50 Ω
With the /N1 option.
Pt50
50 Ω
150 Ω