Withstand voltage, Measurement accuracy, Withstand voltage -9 – Yokogawa µR20000 User Manual

Page 22: Measurement accuracy -9

background image

3-9

SM 04P02B01-01E

Testing

3

Withstand Voltage

Procedures

Perform the test with a withstanding voltage tester, and check whether the results
meet the criteria. Perform the measurement with the power switch turned ON.
Test Duration
The test duration is one minute.

Criteria

Tested Terminals

Applied Voltage

Max Allowable Condition
Leak Current

Between the power and

1.5 kVAC

*1

5 mA

Short the L and N terminals.

earth terminals

Between the input and

1.0 kVAC

1 mA

Short all input terminals.

earth terminals

Between input terminals

1.0 kVAC

1 mA

Short the odd and even
channels of the +/A, −/B,
and b

*2

terminals, and

measure between odd
and even channels.

Between the relay contact

1.5 kVAC

1 mA

With the /A1, /A2, /A3, /A4,

output terminals and

/A5, or /F1 option. Short all

earth terminals

relay contact output terminals.

Between remote control

500 VDC

0.1 mA

With the /R1 option.

input C terminal and
earth terminals

*1 With the /P1 option, the applied voltage is 1.0 kVAC.
*2 With the dot model without the /N2 option, exclude the b terminal. The b terminals are

connected to all channels.

Measurement Accuracy

Overview

Tests the measuring accuracy of the recorder. The measurement accuracy test is
performed in one of the following ways.
• One representative channel

*

for each A/D converter is tested, and the channel-to-

channel error between the representative channel and other channels is tested
(see next item).

*

The pen model uses one A/D converter (A/D 1) for both channels 1 and 2, and another (A/
D 2) for both channels 3 and 4. Choose channel 1 or 2 to test A/D 1, and channel 3 or 4 to
test A/D 2.
The dot model uses a single A/D converter for channels 1 through 24. Choose any one of
channels 1 through 24 to test the A/D converter.

• The measuring accuracy of all channels is tested separately.

Test Instruments

Instrument

Specifications

DC voltage generator

Accuracy: ±0.01% of setting or better

DMM

Accuracy: ±0.01% of rdg or better

Variable resistors

Accuracy: 0.01% + 2 mΩ or better

Recorder Settings

Set the input range of the channels under test to the test range.

3.4 Test Procedures