2 self diagnostic test, Self diagnostic test -2 – Yokogawa µR20000 User Manual

Page 15

background image

3-2

SM 04P02B01-01E

3.2

Self Diagnostic Test

The recorder is provided with complete self diagnostic functions to enhance reliability in
measurement and serviceability.
When you turn ON the power, the recorder will automatically execute the following types
of diagnoses alternately and display the results. After these tests are completed, the
recorder is ready for use.

1.

Main program (Flash ROM) test

2.

Main RAM write/read test

3.

A/D and A/D EEPROM test

4.

Memory acquisition test (write test from Flash)

5.

Mechanical system test

6.

Ethernet module test

The table below shows the results of the self diagnostic tests when a problem is
detected.

Error display

Possible Cause

B3

secondary copy error.

Problem with CPU Assembly

B4

Flash Device error.

B5

Flash SUM error.

B6

Flash Data error.

902 RAM

failure.

910

A/D error.

Problem around the A/D Board PCB or EEPROM

951

EEPROM write error.

921

A/D calibration value error.

Problem with the A/D Board PCB or wrong adjustment data value for

922

A/D calibration is in the wrong order.

the A/D Board PCB.

930

Memory acquisition failure.

Problem with CPU Assembly

940

The ethernet module is down.

Problem with the COMM PCB or CPU Assembly

950

A/D number error.

Problem with the A/D Board PCB (installation of the A/D Board PCB

out of spec)

960 Ribbon

error.

(Dot model) Problem with the ribbon shift mechanism or Dot CPU

Assembly

961 Printer

error.

(Dot model) Abnormal movement of the printing carriage or problem

with the Dot CPU Assembly

962 Plotter

error.

(Pen model) Problem with Pen CPU Assembly

963

Pen 1 error.

(Pen model) Abnormal movement of the Pen Servo Assembly or

964

Pen 2 error.

problem with the Pen CPU Assembly

965

Pen 3 error.

966

Pen 4 error.