Warm up the instrument for thirty minutes before performing the test. Set the ranges
on the recorder, and input the values determined for each range. Check whether the
displayed measured values meet the criteria.

DC voltage generator

To the input
terminals

+

Variable resistors

The resistance of three
lead wires must be equal.

L

N

Criteria

Range

Input Range

Allowed by Test Accuracy Specification

Remarks

20mV

+20 mV

±3 digits

±(0.1% of rdgs + 2digits)

0 mV

±1 digits

−20 mV

±3 digits

60mV +60

mV

±6

digits

0 mV

±1 digits

−60 mV

±6 digits

200mV +200

mV

±3

digits

0 mV

±1 digits

−200 mV

±3 digits

2V +2

V

±3

digits

+1 V

±2 digits

0 V

±1 digits

−1 V

±2 digits

−2

±3 digits

6V +6

V

±6

digits

0 V

±1 digits

−6 V

±6 digits

20V +20

V

±3

digits

0 V

±1 digits

−20 V

±3 digits

50V

+50 V

±5 digits

±(0.1% of rdgs + 3 digits)

0 V

±2 digits

−50 V

±5 digits

Pt100

−200°C: 18.52 Ω

±0.4

°C

±(0.15% of rdgs + 0.3

°C)

0

°C: 100 Ω

±0.2

°C

600

°C: 313.71 Ω

±0.9

°C

TC-T 0

mV

±0.2

°C

EXT

RJC

mode

TC-K 0

mV

±0.3

°C

EXT

RJC

mode

Cu10

−200°C: 1.326 Ω

±1.4

°C

±(0.4% of rdgs + 1.0

°C)

Performed when the /N1 option

(GE) 0

°C: 9.036 Ω

±0.8

°C

is

specified.

300

°C: 20.601 Ω

±1.7

°C

Pt50

−200°C: 8.57 Ω

±0.8

°C

±(0.3% of rdgs + 0.6

°C)

Performed when the /N3 option

0

°C: 50 Ω

±0.4

°C

is

specified.

600

°C: 158.64 Ω

±0.9

°C

3.4 Test Procedures