Dead band (pen model), Pen offset (pen model) – Yokogawa µR20000 User Manual
Page 28
3-15
SM 04P02B01-01E
Testing
3
Dead Band (Pen Model)
● Procedures
Check the accuracy of all pens. Use standard pens (or equivalent).
*
A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.
1.
Turn ON the power switch while pressing the
MENU
and
CH UP
keys. Menu =
A/D Adjust is displayed.
2.
Select Menu = Test, and press the
key.
3.
Choose Test = Dead Band and press the
key.
The dead band measurement pattern is automatically recorded near 0%, 50%,
and 100%. Measure the following portions of the recorded results using a Glass
scale (or equivalent), then confirm that the results meet the criteria.
Criterion: 0.36 mm or less
d
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.
Pen Offset (Pen Model)
● Procedures
Check all pens. Use standard pens (or equivalent).
*
A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.
1.
Turn ON the power switch while pressing the
MENU
and
CH UP
keys. Menu =
A/D Adjust is displayed.
2.
Select Menu = Test, and press the
key.
3.
Choose Test = Dead Band and press the
key.
The dead band measurement pattern is automatically recorded near 0%, 50%,
and 100%.
Measure the following portions of the recorded results using a Glass scale (or
equivalent), then confirm that the results meet the criteria.
Criteria
Item
Criteria
A
2.1 ±1 mm
B
4.2 ±1 mm
C
6.3 ±1 mm
A B C
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.
3.4 Test Procedures