beautypg.com

Appendix 3. setting of av550gspecificcommands, A3.1 av550g-specific commands, A3.2 setting of av550g-specific commands – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 69: A3.1, A3.2, Appendix 3. setting of av550g- specific commands, A-20

background image

A-20

IM 11M12D01-61E

APPENDIX 3. SETTING OF AV550G-

SPECIFIC COMMANDS

Byaccessing the parameters of the transducer block, it is possible to execute commands

specificto the AV550G.

A3.1 AV550G-specific Commands

A3.2 Setting of AV550G-specific Commands

(1) Start Calibration Command
Access the CHx_SEMIAUTO_CAL_START (x = 1 to 8) parameter.

Setting the parameter to 2 (START) allows manual calibration to be performed.

For example, set CH1_SEMIAUTO_CAL_START (189) to 2, to perform manual calibration on the

channel-1 card. (189 is a relative index.)

This function does not operate unless the calibration mode of the AV550G is set to Semi-Auto or Auto.

For details, refer to the AV550G’s user’s manual.

(2) Start Indication Checking Command
Access the CHx_INDICATION_START (x = 1 to 8) parameter.

Setting the parameter to 2 (START) allows indication checking to be performed.

For example, set CH1_INDICATION_START (197) to 2, to perform indication checking on the

channel-1 card. (197 is a relative index.)

This function does not operate unless the indication check mode of the AV550G is set to Semi-Auto or

Auto.

For details, refer to the AV550G’s user’s manual.

Command

Summary

Start Calibration

(CHx_SEMIAUTO_CAL_START, where x is the channel number)

Instructs the AV550G to perform manual calibration on its individual

channel cards.

Start Indication Checking

(CHx_INDICATION_START, where x is the channel number)

Instructs the AV550G to perform indication checking on its individual

channel cards.

Start Blow Back (BLOWBACK_START)

Instructs the AV550G to perform a blow back.

Calibration Gas Pressure Drop (CAL_GAS_PRESS_DROP_SW) Informs the AV550G of a drop in the calibration gas pressure.

Process Gas Alarm (PROCESS_GAS_ALARM_SW)

Informs the AV550G that a process gas alarm has been generated.

NOTE