<5. CONFIGURATION>

5-8

IM 11M12D01-61E

5.6 Block Setting

Set the parameter for function block VFD.

5.6.1 Link Objects

A link object combines the data voluntarily sent by the function block with the VCR. Each AV550G has

40 link objects. A single link object specifies one combination. Each link object has the parameters

listed in Table 5.6. Parameters must be changed together for each VCR because the modifications

made to each parameter may cause inconsistent operation.

Table 5.6 Link Object Parameters

Link objects are not factory-set. Set link objects as shown in Table 5.7.

Table 5.7 Settings of Link Objects (example)

5.6.2 Trend Objects

It is possible to make settings so that a function block automatically transmits trends.

For this, each AV550G has ten trend objects: eight for trends of analog parameters and two for

discrete parameters. For each trend object, specify a single parameter, the trend of which is to be

transmitted.
Each trend object has the parameters listed in Table 5.8. For the first four parameters, setting is

mandatory. Before writing parameter settings to a trend object, parameter WRITE_LOCK of the

resource block must be modified to unlock the write-lock.

T0506.ai

Sub-

index

Parameters

Description

1

LocalIndex

Sets the index of function block parameters to be combined;

set “0” for Trend and Alert.

2

VcrNumber

Sets the index of VCR to be combined. If set to “0”, this link

object is not used.

3

RemoteIndex

Not used in the AV550G. Set to “0”.

5

StaleCountLimit

4

ServiceOperation

Set one of the following. Set only one each for link object

for Alert or Trend.

0: Undefined

2: Publisher

3: Subscriber

6: Alert

7: Trend

Set the maximum number of consecutive stale input values

which may be received before the input status is set to BAD.

To avoid the unnecessary mode transition caused when the

data is not correctly received by subscriber, set this

parameter to “2” or more.

T0507.ai

Index

Link Object #

Settings(example)

30000

AI. OUT VCR#6

1

30001

2

30002

3

30003 to 30039

Not used

4 to 40

Trend VCR#5

Alert VCR#7