Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 56

A-7
IM 11M12D01-61E
26
2026
ERROR_SW_VALUE_D
–
–
Shows that an error has occurred in the AV550G.
27
2027
IN_UNIT
1342
O/S
Denotes the unit of MAO data to be indicated on the AV550G’s
display.
28
2028
IN_DISPLAY_FORMAT
0
O/S
Denotes the decimal places of MAO data to be indicated on
the AV550G’s display.
29
2029
USE_IN_NO
1
O/S
Denotes the IN number of MAO data to be indicated on the
AV550G’s display.
30
2030
PV1_MIN_VALUE
–
–
Denotes the minimum oxygen concentration of averaging A.
31
2031
PV1_MAX_VALUE
–
–
Denotes the maximum oxygen concentration of averaging A.
32
2032
PV1_AVE_VALUE
–
–
Denotes the average oxygen concentration of averaging A.
33
2033
PV1_MIN_DATE
–
–
Denotes the date and time when the minimum oxygen
concentration of averaging A was calculated.
34
2034
PV1_MAX_DATE
–
–
Denotes the date and time when the maximum oxygen
concentration of averaging A was calculated.
35
2035
PV2_MIN_VALUE
–
–
Denotes the minimum oxygen concentration of averaging B.
36
2036
PV2_MAX_VALUE
–
–
Denotes the maximum oxygen concentration of averaging B.
37
2037
PV2_AVE_VALUE
–
–
Denotes the average oxygen concentration of averaging B.
Appendix1-3_2.ai
38
2038
PV2_MIN_DATE
–
–
Denotes the date and time when the minimum oxygen
concentration of averaging B was calculated.
39
2039
PV2_MAX_DATE
–
–
Denotes the date and time when the maximum oxygen
concentration of averaging B was calculated.
40
2040
PV3_MIN_VALUE
–
–
Denotes the minimum oxygen concentration of averaging C.
41
2041
PV3_MAX_VALUE
–
–
Denotes the maximum oxygen concentration of averaging C.
42
2042
PV3_AVE_VALUE
–
–
Denotes the average oxygen concentration of averaging C.
43
2043
PV3_MIN_DATE
–
–
Denotes the date and time when the minimum oxygen
concentration of averaging C was calculated.
44
2044
PV3_MAX_DATE
–
–
Denotes the date and time when the maximum oxygen
concentration of averaging C was calculated.
45
2045
CH1_CELL_VOLT
–
–
Denotes the cell electromotive force of the channel-1 detector.
46
2046
CH1_HEATER_TEMP
–
–
Denotes the heater temperature of the channel-1 detector.
47
2047
CH1_CJ_TEMP
–
–
Denotes the cold junction temperature of the channel-1
detector.
48
2048
CH1_TC_VOLT
–
–
Denotes the thermocouple electromotive force of the
channel-1 detector.
49
2049
CH1_CJ_VOLT
–
–
Denotes the cold junction voltage of the channel-1 detector.
50
2050
CH1_CELL_RESISTANC
E
–
–
Denotes the cell resistance of the channel-1 detector
measured during calibration.
51
2051
CH1_CJ_RESISTANCE
–
–
Denotes the cold junction resistance of the channel-1 detector.
52
2052
CH1_ZERO_CAL_COEFF –
–
Denotes the zero-point correction factor of the channel-1
detector.
53
2053
CH1_SPAN_CAL_COEFF –
–
Denotes the span-point correction factor of the channel-1
detector.
54
2054
CH1_CELL_ROBUSTNES
S
–
–
Denotes the cell service life of the channel-1 detector.
55
2055
CH1_HEATER_ON_TIME
–
–
Denotes the power-on rate of the channel-1 detector’s heater.
56
2056
CH1_RESPONSE_TIME
–
–
Denotes the response time of the channel-1 detector
measured during calibration.
57
2057
CH1_MIN_VALUE
–
–
Denotes the minimum oxygen concentration of channel 1.
58
2058
CH1_MAX_VALUE
–
–
Denotes the maximum oxygen concentration of channel 1.
59
2059
CH1_AVE_VALUE
–
–
Denotes the average oxygen concentration of channel 1.
60
2060
CH1_MIN_DATE
–
–
Denotes the date and time when the minimum oxygen
concentration of channel 1 was calculated.
61
2061
CH1_MAX_DATE
–
–
Denotes the date and time when the maximum oxygen
concentration of channel 1 was calculated.
62
2062
CH2_CELL_VOLT
–
–
Denotes the cell electromotive force of the channel-2 detector.
63
2063
CH2_HEATER_TEMP
–
–
Denotes the heater temperature of the channel-2 detector.
64
2064
CH2_CJ_TEMP
–
–
Denotes the cold junction temperature of the channel-2
detector.