A-18 – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 67
A-18
IM 11M12D01-61E
Example:
To measure 0 to 100%,
Set (1342)*1 in Units Index of XD_SCALE,
Set 100 in EU at 100% of XD_SCALE, and
Set 0 in EU at 0% of XD_SCALE.
(1)-2. Setting the output scale
Example:
To set the output range to 0.00 to 100.00%,
Set (1342)*
1
in Units Index of OUT_SCALE,
Set 100 in EU at 100% of OUT_SCALE,
Set 0 in EU at 0% of OUT_SCALE,
and Set 2 in Decimal Point of OUT_SCALE.
*1: Each unit is expressed using a 4-digit numeric code. Refer to Section 5.6.4, “AI Function Block
Parameters.”
(2) Setting the output mode
(3) Simulation
Perform simulation of each AI function block by setting the desired value and status of the input to the
block.
FA0203.ai
Access the OUT_SCALE parameter.
Set the required unit in Unit Index of OUT_SCALE.
Set the output value corresponding to the upper range limit in
EU at 100% of OUT_SCALE.
Set the output value corresponding to the lower range limit in
EU at 0% of OUT_SCALE.
Set the decimal point position in Decimal Point of OUT_SCALE.
FA0206.ai
Access the L_TYPE parameter.
Set the output mode.
1: Direct
2: Indirect
3: IndirectSQRT
(Sensor output value)
(Linear output value)
(Square root extraction output value)*
1
*1: IndirectSQRT is not used for the AV550G.
FA0207.ai
Access the SIMULATE Value element of SIMULATE and
set the desired input value.
Access the SIMULATE Status element of SIMULATE and
set the desired status code.
Access the En/Disable element of the SIMULATE parameter to
enable simulation.
1: Disabled
2: Active
REMOTE LOOP TEST SWITCH is written to SIM_ENABLE_MSG
(index 1044) parameter of the resource block.