A-18 – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 67

background image

A-18

IM 11M12D01-61E

Example:

To measure 0 to 100%,

Set (1342)*1 in Units Index of XD_SCALE,

Set 100 in EU at 100% of XD_SCALE, and

Set 0 in EU at 0% of XD_SCALE.

(1)-2. Setting the output scale

Example:

To set the output range to 0.00 to 100.00%,

Set (1342)*

1

in Units Index of OUT_SCALE,

Set 100 in EU at 100% of OUT_SCALE,

Set 0 in EU at 0% of OUT_SCALE,

and Set 2 in Decimal Point of OUT_SCALE.

*1: Each unit is expressed using a 4-digit numeric code. Refer to Section 5.6.4, “AI Function Block

Parameters.”

(2) Setting the output mode

(3) Simulation
Perform simulation of each AI function block by setting the desired value and status of the input to the

block.

FA0203.ai

Access the OUT_SCALE parameter.

Set the required unit in Unit Index of OUT_SCALE.

Set the output value corresponding to the upper range limit in

EU at 100% of OUT_SCALE.

Set the output value corresponding to the lower range limit in

EU at 0% of OUT_SCALE.

Set the decimal point position in Decimal Point of OUT_SCALE.

FA0206.ai

Access the L_TYPE parameter.

Set the output mode.

1: Direct

2: Indirect

3: IndirectSQRT

(Sensor output value)

(Linear output value)

(Square root extraction output value)*

1

*1: IndirectSQRT is not used for the AV550G.

FA0207.ai

Access the SIMULATE Value element of SIMULATE and

set the desired input value.

Access the SIMULATE Status element of SIMULATE and

set the desired status code.

Access the En/Disable element of the SIMULATE parameter to

enable simulation.

1: Disabled

2: Active

REMOTE LOOP TEST SWITCH is written to SIM_ENABLE_MSG

(index 1044) parameter of the resource block.