A2.2 Setting and Change of Basic Parameters

This section describes the procedure taken to set and change the parameters for each block.

Obtaining access to each parameter differs depending on the configuration system used. For details,

refer to the instruction manual for each configuration system

Do not turn the power OFF immediately after parameter setting. When the parameters are saved to

the EEPROM, the redundant processing is executed for the improvement of reliability. If the power is

turned OFF within 60 seconds after setting of parameters, changed parameters are not saved and

may return to their original values.

Note 1: Block mode consists of the following four modes that are controlled by the universal parameter

that displays the running condition of each block.

Target: Sets the operating condition of the block.

Actual: Indicates the current operating condition.

Permit: Indicates the operating condition that the block is allowed to take.

Normal: Indicates the operating condition that the block will usually take.

Note 2: The followings are the operating conditions which the individual blocks will take.

Note: Refer to Appendix 1, “List of parameters for each block of the AV550G” for details of the Write Mode for each block.

A2.3 Setting the AI Function Blocks

Each AV550G contains three AI function blocks having independent parameters. Set up the

parameters of each AI block you use, individually as necessary.
The AI block performs the averaging oxygen concentration output calculation.
(1)-1. Setting the calibration range

Access the parameter to be set or changed.

Make setting or change in accordance with each parameter.

Set the Target of block mode (MODE_BLK) back to Auto

(*Note 2)

.

Access the block mode (MODE_BLK) of each block.

FA0201.ai

Set the Target

(*Note 1)

of block mode MODE_BLK) to Auto, Man or O/S

(*Note 2)

according to the Write Mode of the parameter to be set or changed.

IMPORTANT

TA0202.ai

Automatic (Auto)

Manual (Man)

AI Function Block Transducer Block Resource Block DI Function Block

Out of Service (O/S)

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

FA0202.ai

Access the XD_SCALE parameter.

Set the required unit in Unit Index of XD_SCALE.

Set the upper range limit in EU at 100% of XD_SCALE.

Set the lower range limit in EU at 0% of XD_SCALE.

Set the decimal point position in Decimal Point of XD_SCALE.